Pannoramic MIDI II 2.0.5 User’s Guide
5.6.2 Services
◦
Set
Position
,
Objective type
, and
Magnification
settings of the objective.
◦
Use manual camera changer
– Changing the cameras is performed manually.
◦
Prescan objective protection enabled
– If activated, slide area near the edges will not be
scanned to record fast preview image. If inactive, Fast Preview processing becomes
slower..
◦
Cameras –
Activate
BF
and/or
FL
mode for the camera. The software automatically loads
the camera settings upon clicking the
Test and apply camera settings
button.
◦
Preset profiles
are the following:
▪
Calibrated Linear
: Color scheme is defined by calibration (Gamma value is not
included)
▪
sRGB standard
: Color scheme is defined by calibration (standard Gamma value -2.2-
is applied)
▪
sRGB standard Microscope Like
: Color scheme is defined by calibration (standard
Gamma value / White Point: D50)
▪
Legacy
: a Color scheme calibrated for the previous software version is applied
▪
Custom
: parameters are freely customizable as follows:
◦
Color Scheme
▪
Calibrated
: Scheme is defined by the color calibration slide
▪
Legacy
: a Color scheme calibrated for the previous software version is applied
▪
Raw
: the color information of the recorded raw image is applied
◦
White Point
▪
D65
: color temperature according to a standard illuminant D65 white point value is
applied (6500K / daylight)
▪
D50
: color temperature according to a standard illuminant D50 white point value is
applied (5000K / horizontal daylight)
◦
Gamma
▪
Standard
: a value of 2.2 is used
▪
Medium
: a value of 1.8 is used
▪
Off
: a value of 1.0 is used
December 7, 2018 – Rev. 1
3DHISTECH Ltd.
112(144)