MAINTENANCE
70
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
CONFIGURABLE SETTINGS (Continued)
Load Factory Default
The user can load the factory default settings in the
configuration menu.
From the configuration page, pressing CLR key and
key at the same time will bring out the
customization authorization screen.
Enter the password “1111” by pressing
and
once and pressing
11 times. Then press
“SELECT”. The pump will enter the customization
page. Use the up/down
arrow keys to move to
the LOAD DEFAULT option. Use
or
to
remove the “*” sign and choose “SELECT” to save the
changes and return to the configuration menu. Use the
up/down
arrow keys to move to the LOAD
DEFAULT option and press “SELECT”. A
confirmation page will be displayed: “Confirm
Restoration Of Factory Default Parameters?” Pressing
“YES” would reset the pump to factory default
parameters. Press “NO” to cancel the action.