INTRODUCTION
6
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
FEATURES
Flow Rates
The Z-800F Infusion Pump flow rate range is from 1 to
1200 ml/h.
Free Flow Protection
The Z-800F Infusion Pump has a built in free flow clamp
to prevent inadvertent free flow when the set is loaded in
the pump.
The Z-800F Infusion Pump also has an IV Set based anti-
free-flow module to prevent inadvertent free flow when
the set is unloaded from the pump.
WARNING:
Make sure to only use 800F IV Sets
with IV Set based anti-free flow clamp, and make sure to
close the roller clamp before removing administration set
from Z-800F Infusion Pump.
Occlusion Pressure
The Z-800F Infusion Pump provides 14 levels of
adjustable downstream occlusion alarm thresholds
between 4 psi and 30 psi.
Secondary Infusions
Secondary infusions may be delivered at a specified
secondary delivery rate and secondary volume-to-be-
infused (VTBI) independent of the primary infusion
parameters. When secondary VTBI is greater than zero,
the pump will always execute secondary infusion first.
Automatic changeover occurs to the primary infusion
parameters when the secondary infusion is complete. A
Zyno Medical proprietary primary administration set
with a back check valve must be used.
System Configuration
The system configuration mode allows qualified
personnel to customize device settings.
Tamper Resist
The Tamper Resist feature provides a quick, one touch
lockout of the front keypad.
Volume-To-Be-Infused
(VTBI)
The volume-to-be-infused (VTBI) range is from 1 to
9999 ml in 1 ml increments.