MANAGING INFUSION
Z-800F Instructions for Use.
41
P/N 800F-IFU-2602, Rev. O
CLEAR VOLUME INFUSED
1.
From primary or secondary setup state, user may elect
to clear the current volume infused (VINF) parameter.
2.
Press the RUN/STOP key to PAUSE the primary or
secondary infusion during infusion.
3.
Press the PROG key to return to programming page,
press and hold
CLR
key, an audio prompt tone will
sound and a confirmation screen will be displayed.
Press the
YES
key to confirm the action
Press the
NO
key to abort the action.
4.
Upon user confirmation, the current VINF parameter
will be reset to 0 ml.
NOTE: The VINF field will record cumulative infusion
volume infused unless cleared by pressing the
CLR
key.