MAINTENANCE
Z-800F Instructions for Use.
55
P/N 800F-IFU-2602, Rev. O
CONFIGURABLE SETTINGS
Adjusting the Occlusion Pressure Threshold
The pressure sensor detects down-stream occlusion in
the IV sets. A user may specify the occlusion alarm
pressure threshold in configuration screen. The
occlusion alarm threshold can be adjusted between 4
psi and 30 psi
1.
From the configuration menu, Select 1
.
BIOMED OPT
.
2.
In the
BIOMED OPT
menu, highlight the
PRESSURE
option by pressing the
Up/Down
Arrow.
3.
Press
or
key to adjust the pressure.
4.
Press the
SAVE
key
to confirm the set
value. Press the
CANCEL
key
to back out.
Press the
QUIT
key
again to exit the
configuration menu.
WARNING:
The factory default settings
should be used unless qualified clinical personnel
determine that other customized settings are
appropriate and safe.