INFUSION MODE INFORMATION
Z-800F Instructions for Use.
19
P/N 800F-IFU-2602, Rev. O
CONTINUOUS MODE-PRIMARY INFUSION (Continued)
Change Primary VTBI
Use the Up/Down Arrow keys to highlight Pri VTBI. Use
the Data Entry keys to modify the corresponding digits of
Pri VTBI value.
Example
: Programming Pri VTBI = 250 ml
Use the Up/Down Arrow keys to highlight Pri VTBI.
Press the 100 Up data entry key
twice to increase the
100 digit of Pri VTBI to 200.
Press the 10 Up data entry key
five times to increase
the 10 digit of Pri VTBI to 50.
Start Primary Infusion
Verify the displayed infusion parameter entries (Primary
Rate, Primary VTBI). If a clamp is engaged, remove the
clamp. Press
RUN/STOP
key to start the infusion.
Stopping and Resuming a Primary Infusion
Press the
RUN/STOP
key during infusion. The infusion will
be paused. The
PAUSE
screen will be displayed.
From the
PAUSE
state, press the
RUN/STOP
key to resume
infusion.
Titrate Flow Rate or VTBI During a Primary Infusion
From the
PAUSE
screen, press the
PROG
key; the pump
will display primary infusion programming
screen. The user
is then able to modify the primary RATE and primary VTBI
parameters.