MAINTENANCE
60
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
CONFIGURABLE SETTINGS (Continued)
NOTE: Hiding the secondary mode should be completed prior to starting an infusion for
proper administration. When in manual continuous combined infusion, changing HIDE
SEC to YES results in the infusion parameters being zeroed out. When in manual non-
continuous infusion, changing HIDE SEC has no effect on selected infusion.
Configuring Secondary Infusion Option
The Z-800 pump has the option to hide the Secondary
Infusion option so that it is not selectable by the user.
To hide or unhide this option, select the
BIOMED OPT
Menu
. When hiding the secondary infusion option, it
will no longer be selectable by the user.
1.
From the configuration menu, Select 1
. BIOMED
OPT
.
2.
In the
BIOMED OPT
menu, highlight the
HIDE
SEC
option by pressing the
Up/Down Arrow.
3.
Press
or
key to switch between
YES
and
NO
.
4.
Press the
SAVE
key
to confirm the set value.
Press the
CANCEL
key
to back out. Press
the
QUIT
key
again to exit the
CONFIGURATION menu.