MAINTENANCE
56
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
CONFIGURABLE SETTINGS (Continued)
Adjusting the Air-In-Line Alarm Threshold
The Z-800F’s Air-In-Line sensor detects an air bubble
in the administration set tubing. Air-In-Line alarm
threshold can be adjusted in the configuration menu.
Alarm threshold can be adjusted to air bubble size from
a volume of 2 micro-liters to 280 micro-liters. The Z-
800F’s Air-In-Line Detector alarm threshold can be
adjusted by the user.
1.
From the configuration menu, Select 1
. BIOMED
OPT
.
2.
In the
BIOMED OPT
menu, highlight the
AIR
option by pressing the
Up/Down Arrow.
3.
Press
or
key to adjust the value.
4.
Press the
SAVE
key
to confirm the set value.
Press the
CANCEL
key
to back out. Press
the
QUIT
key
again to exit the configuration
menu.
WARNING:
The factory default settings
should be used unless qualified clinical personnel
determine that other customized settings are
appropriate and safe.