INFUSION MODE INFORMATION
24
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
CONTINUOUS MODE-SECONDARY INFUSION (Cont.)
Start Secondary Infusion
Verify displayed infusion parameter entries (Primary Rate,
Primary VTBI, Secondary Rate, and Secondary VTBI). If
a clamp is engaged, remove the clamp. Press
RUN/STOP
key to start the infusion.
NOTE:
Secondary infusion will be executed before
primary infusion.
Secondary Infusion Execution
The two Infusion Status Indicator LEDs will
illuminate in green and toggle once per second.
The secondary infusion execution display contains the
following contents:
Status Bar: The Status Bar displays current operating mode;
current pump state; and volume infused.
SEC
: Secondary infusion
RUN
: Pump RUN state
TL
: Time Left in hours: minutes (HH:MM)
format for the current infusion.
Infusion Parameters: Current infusion parameters
RATE
field displays secondary flow rate.
VTBI
field counts down to show remaining
VTBI.
VINF
field counts up to show volume infused.
Upon completion of secondary infusion, a switchover
audio alert sounds 4 short beeps. The Main Display
will switch-over to
PRI (primary) State.
WARNING: Do not over-program VTBI.
Program the actual amount of the fluid in the IV bag.