ALARMS AND TROUBLESHOOTING
48
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
ALARMS
Alarm
Meaning Secondary
Alarm/
Warning Light (optional)
Response
Air-in-
line
Air has been detected
in the set during an
infusion. Infusion
stopped.
Continuously flashes (blue
or red) notification until
user acknowledgement.
1.
Acknowledge the alarm by
pressing
RUN/STOP key
.
2.
The infusion is paused.
3.
Clear air from line and
press the
RUN/STOP key
again to resume infusion.
Door
Open
Pump door is opened
during an infusion.
The infusion is
stopped.
Continuously flashes (blue
or red) notification until
user acknowledgement.
1.
Acknowledge the alarm by
pressing the
RUN/STOP
key
.
2.
The infusion is paused.
3.
Close the pump door. Press
the
RUN/STOP key
again
to resume infusion.
Occlusion Increased
back
pressure is sensed
while infusing.
Infusion is stopped.
Continuously flashes (blue
or red) notification until
user acknowledgement.
1.
Resolve the cause of the
occlusion.
2.
The infusion will resume
when pressure is reduced to
below the alarm threshold.
No Drip
The drip sensor does
not detect a fluid drop
in the drip chamber
(continues for 8mL,
then alarms).
Continuously flashes (blue
or red) notification until
user acknowledgement.
1.
Acknowledge the alarm by
pressing
RUN/STOP key
.
2.
The infusion is paused.
3.
Verify that the fluid
container is not empty.
Press the
RUN/STOP key
again to resume infusion.
Battery
Empty
The Z-800F pump is
operating on battery
power and battery is
too low for pump
operation.
Continuously flashes (blue
or red) notification until
user acknowledgement.
1.
Plug the power cord into an
AC power outlet.
2.
Acknowledge the alarm by
pressing
RUN/STOP key
.
3.
The infusion is paused.
Press
RUN/STOP key
again
to resume infusion.