MAINTENANCE
Z-800F Instructions for Use.
61
P/N 800F-IFU-2602, Rev. O
CONFIGURABLE SETTINGS (Continued)
Enable/Disable Clamp Sensor Option
The user may enable or disable the clamp sensor in the
configuration menu. When enabled, the clamp sensor
will detect the absence of the IV set based free flow
protection clamp, alarm and stop the pump when the
user tries to start an infusion.
1.
From the configuration menu, Select 1
. BIOMED
OPT
.
2.
In the
BIOMED OPT
menu, highlight the
CLAMP SENSOR
option by pressing the
Up/Down Arrow.
3.
Press
or
key will bring out the
customization authorization screen.
4.
Enter the password “ 1111” by pressing
and
once and pressing
11 times. Then press
“SELECT.” The pump will return to the
BIOMED OPT
menu.
5.
Use
or
to switch between YES (Enable)
and No (Disable).
6.
Press the
SAVE
key
to confirm the set value.
Press the
CANCEL
key
to back out. Press
the
QUIT
key
again to exit the
CONFIGURATION menu.
WARNING:
The factory default settings
should be used unless qualified clinical personnel
determine that other customized settings are
appropriate and safe.