INFUSION MODE INFORMATION
Z-800F Instructions for Use.
21
P/N 800F-IFU-2602, Rev. O
LOADING SECONDARY ADMINISTRATION SET
1.
Program and start the primary infusion using a check-
valve primary administration set, as previously
described.
2.
Open a secondary administration set package, remove
the set and close the set roller clamp.
3.
Insert set spike into a prepared fluid container and hang
secondary container, following accepted clinical
procedures.
4.
Fill the drip chamber to 1/3 full.
5.
Open a secondary set and prime the set. Close the set
roller clamp.
6.
Attach a secondary set to the upper injection site on the
primary set.
7.
Ensure that no air bubbles are in the line.
8.
Hang the secondary fluid container at least 8 inches
above the primary solution container.
WARNING: Make sure to close roller clamp before
removing administration set from Z-800F pump.
WARNING: The bottom of the secondary
solution should be at least 19” above the top of the
pump.
WARNING: A minimum height differential
between primary and secondary solutions of eight inches
is essential for the safe operation of a primary/secondary
infusion.