MAINTENANCE
Z-800F Instructions for Use.
53
P/N 800F-IFU-2602, Rev. O
MAINTENANCE
SPECIFICATIONS
Pumping Mechanism:
Linear Peristaltic
Flow Rate:
1 - 1200 ml/h in 1 ml increments
Priming Rate:
600 ml/h
Volume to Be Infused:
1 - 9,999 ml in 1 ml increments
Accuracy*:
± 5 %
Time Memory is maintained:
Permanent for previous infusion parameters.
Air-in-Line Detector:
Ultrasonic, 14-level configurable
Occlusion Detector:
Pressure sensor. 14-level configurable
Electrical Standard:
Class I, Type BF
Electrical Safety
Complies with: EN 60601-2-24:1998; EN 60601-
1:2006/AC:2010; EN 60601-1-2:2007/AC:2010; EN
60601-1-6:2010; EN 60601-1-8:2007; EN 60601-1-
11:2010, and EN 62366:2008.
Power Requirements:
100 – 240 VAC, 50-60 Hz
Power Consumption:
25 VA (at maximum flow rate)
Internal Battery:
Rechargeable Nickel-Metal-Hydride
9.6V; 4.5 Amp-hr
(Replaceable by qualified service personnel only)
Battery Life at 125 ml/h:
8 hours
Battery Charging:
Automatic when pump plugged into an AC power source
Pump Housing:
Cast Aluminum & Sheet Metal
Weight:
3.2 kilograms, 7.0 lbs
Dimensions:
8.6”H x 5.7”W x 5.3”D
Standards:
Manufactured in a facility certified with ISO 13485: 2003
*The device is calibrated within flow rate accuracy specification under the following conditions:
Test fluid is distilled water at room temperature (21+/-1
C) at ambient air pressure;
Head height is set according to instructions in the IFU;
Test set is Zyno proprietary set at specifications;
Programmed volume during test is no less than 10 mL.