GETTING STARTED
Z-800F Instructions for Use.
15
P/N 800F-IFU-2602, Rev. O
PREPARING
AN INFUSION
Powering On the System
1.
Connect the Z-800F Infusion Pump Power Cord to an
AC power source.
2.
Press and hold the
ON/OFF
key on the Z-800F Infusion
Pump for 2 seconds.
3.
System self-test begins:
The diagnostics test causes all LED display
segments and Status indicator lights to illuminate
briefly.
The Power Indicator illuminates.
An Audio tone sounds.
4.
The Main Display shows the Zyno Medical Logo, Pump
Serial Number, Software Version and pump safety
configurations during the system power-on self-test.
5.
Upon completion of system power-on self-test, a New
Infusion prompt screen will be displayed prompting the
user to choose either resume a previously interrupted
infusion or start with a new infusion.
6.
If the user elects to resume the previously interrupted
infusion, the pump will enter into the Programming
screen of the previous selected infusion mode and with
the previous infusion parameters populated.
7.
If the user elects to start a new infusion, the pump will
prompt user to main screen (see right image). For
manual programming, the user should select
PROGRAM. After selection, the pump will prompt user
to select infusion mode. Upon selection, the pump will
enter into the Infusion Programming screen of the
selected Infusion Mode with all infusion parameters
reset to zero. The user may choose one of the following
infusion modes:
Continuous Mode with Rate/VTBI parameters
Continuous Mode with Time/VTBI parameters
TPN Auto Ramp Mode
10 Step Mode Rate/VTBI
10 Step Mode Time/VTBI
Intermittent
Blood Infusion