MAINTENANCE
Z-800F Instructions for Use.
59
P/N 800F-IFU-2602, Rev. O
CONFIGURABLE SETTINGS (Continued)
Enable/Disable Drip Sensor Option
The Z-800F pump has an optional external drip sensor,
which may be used to detect a fluid container empty
event. The user may enable or disable the drip sensor in
the configuration menu. To enable or disable this
option, refer to “
Customize the BIOMED OPT Menu
”
page in this section. When enabled, the drip sensor will
detect the absence of drips in the drip chamber, continue
to infuse 8mL, then alarm and stop the pump.
1.
From the configuration menu, Select (1)
.
BIOMED OPT
.
2.
In the
BIOMED OPT
menu, highlight the
DRIP
SENSOR
option by pressing the
Up/Down
Arrow.
3.
Press
or
key to switch between
YES
and
NO
.
4.
Press the
SAVE
key
to confirm the set value.
Press the
CANCEL
key
to back out. Press
the
QUIT
key
again to exit the
CONFIGURATION menu.