CMM
61211-1310402
CONTURA
4-7
Example of a reference temperature for measurement of 20°C:
Temperature
Change
Status
Action
Per day
18.6 - 20.0°C
1.4 K
OK
20.0 - 21.5 °C
1.5 K
OK
18.3 - 20.0 °C
1.7 K
Not OK
Wait for 48 hours until the newly
set temperature is constant.
Per hour
19.0 - 20.0°C
1 K
OK
20.0 - 20.9 °C
0.9 K
OK
18.8 - 20.0 °C
1.2 K
Not OK
Wait for 48 hours until the newly
set temperature is constant.
Per meter
19.0 - 20.0°C
1 K
OK
20.0 - 20.9 °C
0.9 K
OK
18.8 - 20.0 °C
1.2 K
Not OK
Wait for 48 hours until the newly
set temperature is constant.
Содержание CONTURA
Страница 46: ...Components and function 3 10 61211 1310402 CONTURA Attaching the holders to the profile rail 6 9...
Страница 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Страница 62: ...CMM 4 8 61211 1310402 CONTURA...
Страница 68: ...Installation 5 6 61211 1310402 CONTURA...
Страница 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Страница 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Страница 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Страница 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Страница 130: ...Service features 8 8 61211 1310402 CONTURA...
Страница 131: ...9 61211 1310402 CONTURA 9 1 Maintenance and care This chapter contains Maintenance 9 2 Care 9 4 Maintenance and care...
Страница 150: ...2Glossary 2 61211 1310402 CONTURA...
Страница 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Страница 157: ......
Страница 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...