Special measures
8-6
61211-1310402
CONTURA
In case of air conditioning failure
In the event of a failure of the air-conditioning unit in the inspection
room, you must proceed as follows after its reparation:
1
Slowly adjust the temperature to the nominal value.
The temperature gradient should be about 1 K per hour. Larger gra-
dients may cause material deformations, thus leading to permanent
deformation of the CMM guideways.
2
Once the nominal temperature has been reached, wait at least for
one day before starting the measuring operation.
This is important to allow possible deformations to disappear and to
ensure an optimal measuring operation.
Ideally, you should also check the precision of the CMM using a suitable
standard, e.g. CMM check.
Содержание CONTURA
Страница 46: ...Components and function 3 10 61211 1310402 CONTURA Attaching the holders to the profile rail 6 9...
Страница 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Страница 62: ...CMM 4 8 61211 1310402 CONTURA...
Страница 68: ...Installation 5 6 61211 1310402 CONTURA...
Страница 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Страница 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Страница 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Страница 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Страница 130: ...Service features 8 8 61211 1310402 CONTURA...
Страница 131: ...9 61211 1310402 CONTURA 9 1 Maintenance and care This chapter contains Maintenance 9 2 Care 9 4 Maintenance and care...
Страница 150: ...2Glossary 2 61211 1310402 CONTURA...
Страница 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Страница 157: ......
Страница 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...