Preparation for measuring operation
61211-1310402
CONTURA
7-11
Starting the automatic run
The homing run is started by the measuring software. Please refer to the
operating instructions for the measuring software.
Park position
After homing, the probe moves to the park position. The park position
coordinates are defined in the measuring software. The coordinates
must be inside the measuring volume.
Example
of park position: X = 30, Y = -30, Z = -30.
NOTE
The position of the reference sphere has to be redetermined in the fol-
lowing cases after a homing run:
Case 1: Redetermination is necessary if a stylus system must be requali-
fied. This is required, for example, after supplementing or replacing a
stylus and after a collision.
Case 2: Redetermination is necessary if the position of the changer rack
or
holder has changed.
Stylus system qualification
What you should know!
Why is qualification necessary?
By means of qualification, the stylus vector and diameter are deter-
mined. This data is required to determine the coordinates of the probed
point.
The reference measurement has to be carried out before determining
the stylus vector.
In which cases is qualification necessary?
Qualification is necessary in the following cases:
–
After a collision
–
When the ambient temperature varies considerably and if the stylus
system consists of components having different expansion coeffi-
cients.
–
When inserting a new stylus into the stylus system.
–
When mounting a new probe.
What should be observed during qualification?
Take the following steps during qualification:
Содержание CONTURA
Страница 46: ...Components and function 3 10 61211 1310402 CONTURA Attaching the holders to the profile rail 6 9...
Страница 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Страница 62: ...CMM 4 8 61211 1310402 CONTURA...
Страница 68: ...Installation 5 6 61211 1310402 CONTURA...
Страница 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Страница 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Страница 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Страница 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Страница 130: ...Service features 8 8 61211 1310402 CONTURA...
Страница 131: ...9 61211 1310402 CONTURA 9 1 Maintenance and care This chapter contains Maintenance 9 2 Care 9 4 Maintenance and care...
Страница 150: ...2Glossary 2 61211 1310402 CONTURA...
Страница 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Страница 157: ......
Страница 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...