2.4.1.5. EMO button
The emergency off button (EMO button) is
located on the plinth.
The EMO button is to be pressed in an emergency
to cut off power to all devices connected to the
EMO box.
It must always be readily accessible and operable.
2.4.1.6. Main shut-off valves
The user is responsible for the installation of main shut-off valves at the site of installation. The
following main shut-off valves are required:
•
water supply
•
water runback
•
nitrogen supply
•
compressed air supply
The main shut-off valves have to be easily accessible. They must close off the connections to the
corresponding media when needed. The main shut-off valves have to be lockable in their OFF
position in order to prevent accidental re-activation.
As the user is responsible for installing the main shut-off valves, he/she should also provide in-
structions how to operate the main shut-off valves properly.