Click
Save
to save the adjustments.
8
Repeat steps 5 to 7 with the next smaller probe
current.
9
Adjust all smaller probe currents with this proce-
dure.
10 Select the 30kV:50 pA reference current to
check the original position of the crosshairs. If
necessary move the crosshairs to the original
position.
CAUTION
Danger of damaging the specimen
Ensure not to damage the specimen while working with high currents.
11 Repeat steps 6 to 7 with the next higher probe
current.
12 Adjust all higher probe currents with this proce-
dure.
Now, the adjustment is complete.