4-46
IM AQ6376-01EN
When SMPL ENABLE Is Selected
The minimum necessary pulse width of the external trigger input signal varies according
to the sensitivity setting.
If the time that the external trigger input signal level is low is less than this pulse width,
the instrument cannot perform the sweep.
Sensitivity Name and Minimum Pulse Width
Sensitivity Setting
Minimum Pulse Width
NORM/HOLD
50 ms
NORM/AUTO
50 ms
NORMAL
50 ms
MID
50 ms
HIGH1/CHOP
100 ms
HIGH2/CHOP
250 ms
HIGH3/CHOP
2050 ms
4.16 External Trigger Measurement