6-65
IM AQ6376-01EN
Analysis
1
2
3
4
5
6
7
8
9
10
11
App
Index
Executing and Stopping Data Logging
1.
Press
ADVANCE
and then the
DATA LOGGING
soft key. The soft key menu for
data logging appears.
2.
Press the
START
soft key. A confirmation message for deleting the existing
logging data and the EXECUTE and CANCEL soft keys appear.
3.
Press the
EXECUTE
soft key. The existing logging data is deleted, and a new
data logging session begins.
When the specified number of measurements is reached, data logging
automatically stops.
If you press the
STOP
soft key while data logging is in progress, data logging will
be stopped.
If you do not want to delete the existing logging data, press the
CANCEL
soft key.
The previous soft key menu will return.
Note
• When TRACE LOGGING is ON, the AQ6376 checks the amount of free space in the
temporary trace data storage space before starting to log. If there is insufficient free memory
space, a warning will appear.
WARNING 151: Disk space is not enough for logging
If this happens, reduce the number of waveform data values (SAMPLING POINT) or reduce
the logging time to decrease the waveform data size.
• If there is no existing logging data, the confirmation message for deleting the data will not
appear, and data logging will begin immediately.
• Only the STOP soft key is valid while data logging is in progress.
If you press a soft key other than the STOP soft key, a confirmation message for stopping
the data logging will appear. If you do not want to stop data logging, press the NO soft key.
6.15 Analysis Data Logging