5-13
IM AQ6376-01EN
W
aveform Display
1
2
3
4
5
6
7
8
9
10
11
App
Index
5.5 Displaying Calculated Waveforms
Procedure
Selecting Traces for Calculation
1.
Press
TRACE
. The soft key menu for trace settings appears.
2.
Press the
ACTIVE TRACE
soft key. Traces A through G appear in soft keys.
3.
Press the soft key corresponding to a trace that is available for trace-to-trace
calculation (C, F, or G).
The calculated results are displayed in the selected trace.
The items that can be calculated differ depending on the trace selected.
4.
Press the
VIEW
soft key of the selected trace and select
DISP
. Each time you
press the soft key
DISP
switches to
BLANK
, or viceversa.
Selecting a Calculation
5.
Press the
CALCULATE
soft key. The LOG and Linear selection menus are
displayed.
6.
For log calculations, press the
LOG MATH
soft key. For linear calculations, press
the
LIN MATH
soft key. The soft key menu showing the choices for calculation
appears.
7.
Press the soft key corresponding to the desired calculation. The calculation
executes.
Note
• Trace-to-trace calculations can only be performed on C, F, or G. If the active trace is set to
something other than trace C, F, or G, the CALCULATE key is disabled.
• If a trace targeted for calculation is remeasured and the center wavelength and
measurement span of the trace being measured is changed, it is recalculated and
redisplayed.
• If the measurement condition (resolution) of the trace targeted for calculation does not
match, a warning is displayed after calculation.