1-10
IM AQ6376-01EN
1.3 Waveform Display
Single Search
• PEAK SEARCH
A peak search (a search for the maximum value) is performed on the marker trace
waveform.
An advanced marker is displayed at the peak point, and the marker value is displayed
in the data area.
If the peak level is off the screen, its marker is displayed at the top or bottom edge of
the screen. Even in this situation, the correct marker value is displayed.
After the measurement has been completed, you can use the rotary knob to move
the moving marker.
You can use the arrow keys to scroll through the data area.
• BOTTOM SEARCH
A bottom search (a search for the minimum value) is performed on the marker trace
waveform.
An advanced marker is displayed at the bottom point, and the marker value is
displayed in the data area.
If the bottom is off the screen, its marker is displayed at the top or bottom edge of the
screen. Even in this situation, the correct marker value is displayed.
After the measurement has been completed, you can use the rotary knob to move the
moving marker. You can use the arrow keys to scroll through the data area.
• NEXT LEVEL SEARCH
On the marker trace waveform, the advanced marker at the peak or bottom is set to
the next peak (local maximum value) or bottom (local minimum value).
If such peak or bottom is not found, a warning appears.
WARNING 103: No data in active trace
• NEXT SEARCH RIGHT
On the marker trace waveform, the advanced marker at the peak or bottom is set to
the peak (local maximum value) or bottom (local minimum value) on its right.
If such peak or bottom is not found, a warning appears.
WARNING 103: No data in active trace
• NEXT SEARCH LEFT
On the marker trace waveform, the advanced marker at the peak or bottom is set to
the peak (local maximum value) or bottom (local minimum value) on its left.
If such peak or bottom is not found, a warning appears.
WARNING 103: No data in active trace
Displaying Wavelength Difference and Level Difference
This function places a fixed marker and measures the wavelength difference and level
difference compared to a moving marker.
Note
For details, see the explanation in section 5.9, “Displaying Markers.”
Displaying Line Markers
When line markers are displayed, the marker values are shown in the upper left part of
the waveform area.
When both wavelength line markers 1 and 2 are displayed, or both level line markers
3 and 4 are displayed, the wavelength difference (L2-L1) or level difference (L4-L3) is
shown below the marker values.
Note
For details, see the explanation in section 5.9, “Displaying Markers.”