6.8 Optical Amp Gain and NF Measurement (Except
Wavenumber Mode)
Procedure
Measurement can be performed of the optical amp gain and noise figure from the
measured waveform of the signal light going into the optical amp, and the measured
waveform of the output light leaving the optical amp. (Except Wavenumber Mode)
1.
Press
ANALYSIS
. The soft key menu for analyzing measured waveforms
appears.
2.
Press the
ANALYSIS 2
soft key. The analysis function selection menu is
displayed.
3.
Press the
EDFA-NF
soft key. Analysis is performed, and the results are displayed
in a list. The analysis results display screen is switched with the
SWITCH
DISPLAY
soft key.
Note
When zooming a waveform, if you click in the list of analysis results, the waveform of the
selected channel is displayed in the center of the waveform screen.