7-21
IM AQ6376-01EN
Saving/Loading Data
1
2
3
4
5
6
7
8
9
10
11
App
Index
Waveform Data
Measures waveform data are stored as wavelength values and level values for the
number of measurement sampling points.
A waveform measured in frequency mode is also stored as wavelength values. Level
values are stored as log values if the vertical axis scale is LOG, and as linear values if
the vertical axis scale is linear.
(For LOG Scale)
[TRACE DATA]
Header indicating the start of the trace data
****.****, ±***.***(CR)(LF)
Wavelength value and level value (LOG) of first point
****.****, ±***.***(CR)(LF)
Wavelength value and level value (LOG) of second point
:
****.****, ±***.***(CR)(LF)
Wavelength value and level value (LOG) of final point
(For LINEAR Scale)
[TRACE DATA]
Header indicating the start of the trace data
****.****,*.***E±***(CR)(LF) Wavelength value and level value (LINEAR) of first point
****.****,*.***E±***(CR)(LF) Wavelength value and level value (LINEAR) of second point
:
****.****,*.***E±***(CR)(LF) Wavelength value and level value (LINEAR) of final point
7.3 Saving/Loading Displayed Data