6-42
IM AQ6376-01EN
6.12 Go/No-Go Judgment (Template)
Procedure
This function compares preset reference data (template data) with a measured
waveform, and makes a Go/No-Go judgment.
Creating Template Data on the Instrument
1.
Press
ADVANCE
followed by the
TEMPLATE
soft key.
2.
Press the
TEMPLATE EDIT
soft key. The template creation screen is displayed.
3.
Press the
LINE SELECT
soft key.
4.
Press the soft key corresponding to the type of template you wish to create.
UPPER LINE: Upper limit line
LOWER LINE: Lower limit line
TARGET LINE: Target value line
5.
Press the
MODE ABS/REL
soft key to select either ABS (absolute) or REL (relative)
as the type (template data types).
6.
Press the
EXTRAPOL TYPE
soft key to select the extrapolation method.
TYPE A: Extrapolation type A
TYPE B: Extrapolation type B
NONE:
No extrapolation
7.
To edit template data, move the cursor to the location of the data to be edited
using the rotary knob or arrow keys, then press the
VALUE EDIT
soft key. Enter a
value using the numeric key pad, rotary knob, or arrow keys.
8.
To add template data, press the
INSERT
soft key. The data at the cursor location
is added. Edit the value using the procedure in step 7 and set it as new data.
9.
To delete template data, move the cursor to the location of the data to be deleted
using the rotary knob or arrow keys, then press the
DELETE
soft key. To delete all
template data points, press the
ALL DELETE
soft key.
Note
• For information about extrapolation methods, see the explanation.
• The same wavelength/level data prior to insertion are inserted as data in the points that
were inserted using the INSERT soft key.
• When the target line is turned OFF with the TEMPLATE DISPLAY soft key, if you edit the
template data at the target line, the target line TEMPLATE DISPLAY turns ON.