7. How to use the communication interface
meastemp
Purpose
Measured temperature request
Description
BF500 sends measured temperature data after an echo back when it receives this
command from the host. Contents of the measured temperature data are the
same as those on the operation panel. The echo back and the transmission data
are as follows:
meastemp <CR> <LF>
n <CR> <LF>
(n is measured temperature data)
01>
acctime
Purpose
Request for accumulated operation time.
Description
BF500 sends accumulated operation time data after an echo back when it
receives this command from the host. The echo back and the transmission data
are as follows:
acctime <CR> <LF>
n <CR> <LF>
(n is accumulated operation time data)
01>
beepmode
Purpose
Confirmation of the beep mode status.
Description
BF500 allows setting ON/OFF of an alarm beep when the integrated trouble
detection mechanism detects a trouble. The state of this setting (ON or OFF) is
called beep mode state. BF500 sends the present beep mode state after an echo
back when it receives this command from the host. The echo back and the
transmission data are as follows:
beepmode <CR> <LF>
n <CR> <LF>
n
:
0
:
Beep mode OFF state
01>
1
:
Beep mode ON state
beepmode_n (The underbar “_” means a space)
Purpose
Setting of the beep mode state.
Parameter
The parameter n has the following settings:
0
:
Beep mode OFF state
1
:
Beep mode ON state
Description
BF500 sends an echo back and the ID prompt and changes the present beep
mode state according to the n setting. The echo back and the transmission data
are as follows:
Beepmode_n <CR> <LF>
01>
(The underbar “_” means a space.)