15
4. Operating procedures (For BF400/500/600)
Operations after turning the power switch ON
When operation is ready start operation following the procedures below.
(
Note
:
Model BF600 does not support indication of a temperature with a decimal point
)
Display after operation and operating procedures
Description
1
Turn the power
switch ON
When power is turned on, the main
display indicates the present measured
temeperature and the sub display
indicates the flashing set temperature in
the previous session.
The [STAND BY] lamp comes on.
。
The [FIXED-TEMP] lamp flashes.
2
Example
:
Fixed-temperature operation at
56.0
℃
Press the [
▼
] or
the [
▲
] key.
Press the [
▼
] or the [
▲
] key until the
temperature flahses in the sub
display reaches the target
temperature.
3
Press the
[ENTER] key.
When the target temperature is reached,
press the [ENTER] key.
The set temperature in the sub display
changes its status from flashing to
staying lit and fixed-temperature
operation starts towards the set
temperature.
The [STAND BY] lamp goes off and the
[FIXED TEMP] lamp changes its status
from flashing to staying lit.
When the set temperature is higher than
the present measured temeperature, the
[HEAT] lamp comes on and heating
starts.
Now procedures for the
fixed-temperature operation have been
completed.
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU
Example:Measured temperature: 25.0℃
Example:Set temperature:37.0℃
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU