18
4. Operating procedures
Procedures for program operation
Display after operation and operating procedures
Description
5
Example
:
To start operation after 2 hours
and 30 minutes
Press the [
▼
] or
the [
▲
] key.
Press the [▼] or the [▲] key to flash the
time to start the program you have called
for on the main display.
*
You can set a wait time in the range from
one minute to 999 hours. Note, however,
that a time longer than one hour cannot
be set in the unit of minutes only (for
example, 80 minutes).
Example:99 hours 59 minutes
:
100 hours
:
6
Press the
[ENTER] key.
Press the [ENTER] key.
Indication in the main display changes to
the present measured temeperature.
Indication in the sub display changes to
the remaining time before starting
operation.
The [STAND BY] lamp changes to
flashing to indicate that the unit is in the
stand by state.
The [PROGRAM] operation lamp
changes its status from flashing to
staying lit.
When the operation time comes, the
[STAND BY] lamp goes off and the
program operation starts.
At the operation start time, the sub
display indicates the information set in
the previous session.
*
The example shows that the segment 1
of the program number 2 is in operation
and the remaining time until the end of
the 1
st
segment lamp or the soak interval
is 10 minutes.
*
During the program operation, you can
select information in the sub display from
among: “Remaining time before
completion of the execution segment”,
“Set temperature” and “Number of the
program being executed”. (See P.26 “To
change the sub display indication”.)
<
To change the sub display indication
>
Press the [MODE]
key to switch to
the sub display
switch mode.
7
<
Program
operation ends
>
Temperature control stops.
The main display indicates the present
measured temperature.
The sub display indicates flashing
.
The [HEAT] lamp goes off.
The [PROGRAM] lamp stays lit.
Set temperature
being executed
Number of
program being
executed
Remaining time before
the end of segment
being executed
Remaining time is
set
Execution program
indication is set
Set temperature
indication is set
Number of
segment being
executed
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU
MEASURED TEMP
MEASURED TEMP
MEASURED TEMP