35
5.How to use the function menu (MODE key)
How to set and cancel communication lock out
The function is used to select a process to hand down the controller control to the host computer
(communication lock out cancelled) in response to a communication request from the host
computer connected via a communication interface, or to reject such request (communication lock
out enabled). Communication will be forced to shut off when communication lock out is carried out
while communication is active with the host computer.
Display after operation and operating procedures
Description
1
Press the [MODE]
key.
Press the [MODE] key.
When pressed during operation, the
indication changes to the on in the sub
display switching mode and when
pressed during stand-by or at the end of
a program, indication in the preset
temperature registration mode will
appear.
2
Press the [
▼
] or
the [
▲
] key.
Press the [▼] or the [▲] key several
times.
The main display indicates flashing
[com.L] (abbreviation of communication
lock out).
The sub display indicates the latest
settngs.
3
Press the [ENTER]
key.
Press the [ENTER] key.
。
The mode changes to the mode for
setting/cancelling the communication
lock out function.
Indication in the sub display changes to
[com.L].
4
Press the [
▼
] or
the [
▲
] key.
Press the [▼] or the [▲] key.
The main display indicates flashing
[OFF] and [ON] alternately.
5
〈
Indication immediately before the MODE key
is pressed
〉
Press the [ENTER]
key.
Press the [ENTER] key.
Indication will return to the status
immediately before the [MODE] key is
pressed.
or
Example of
indication of the
preset temperature
registration mode
Example of
indication of the
sub display
switching mode
Communication
lock out is
disabled
Communication
lock out enabled