37
5.How to use the function menu (MODE key)
Program configuration
Description
3. Hold
The program will maintain the target temperature
setting (ramp level) if you set
as the soak
time. The program input mode will end.
flashes at somewhere between 0 hour 0
minute and 999 hours when selecting a soak time.
(See the indication example when hold is selected
on the right)
Note
:This function is different from the hold
function of the
MODE
(status at the time of
entering is maintained) and shall not be
mixed up.
Contents of the repeat command
■
Configuration of the repeat command
The repeat command is to repeat a group of segments
many times and described with a “repeat start
segment” and “number of repetitions”.
1. Repeat start segment
The mode is for inputting segments and is used to
enter a number of segment at which you want to
start repetitions. As a result, the part between the
starting segment and the segment immediately
before inputting is operated repeatedly. (See the
sample indication of repeat input on the right
)
2. Number of repeats
You can set the number of repeats freely from 1 to
999 times.
Note
: You cannot repeat a segment which is set
after a repeat command in that repeat
itself.
Note:
You cannot make setting that overlaps a
repeat command with the setting of a
different repeat command (nesting) or
crosses two repeat commands over.
Note
:
Do not set “0” as the number of repeats.
Program may not be executed properly.
〈
Indication example when hold is selected
〉
〈
Indication of the number of repeat
setting mode
〉
〈
Indication of the repeat start
segment number setting mode
〉
〈
Indication of the repeat command
input mode
〉
〈
Indication example of the segment
input mode
〉
Press [▼] [▲].
Press [▼] [▲].
(
Hold
)
(
999hrs.
)
(0 hr. 0 min)
Press [▼] [▲].
Press [ENTER].
Select with [▼] [▲] and press
[ENTER]
Select with [▼] [▲] and press
[ENTER]
(Starting
segment)
(Number of
repeats)