4. Operating procedures
When you want to switch to a different operation mode
When you want to the program operation to a different program
You can change the “number of the program to execute” and “Operation start wait time” during the
program operation (during execution, stand-by and at the completion of operation). Follow the
procedures below when you want to change to another program operation during a program
operation or to change the operation start wait time.
Display after operation and operating procedures
Description
1
Press the [
▼
] or
the [
▲
] key.
Press the [▼] or the [▲] key.
The mode changes to the selection mode
of a program to execute and the main
display indicates the flashing number of
the program being executed
(Example: program number 2).
The sub display indicates
.
2
Press the [
▼
] or
the [
▲
] key.
When two or more execution programs
exist, press the [▼] or the [▲] key to flash
numbers of executable programs in turn
in the main display.
Make the number of the program to
execute
(
(Example: meaning
program number 3) flash in the main
display.
*
Skip step 2 and move to step 3 directly if
you want to change “Operation start wait
time” only.
3
Press the
[ENTER] key.
Press the [ENTER] key.
The program currently being executed
stops and the [STAND BY] lamp comes
on.
The status of the [PROGRAM] lamp
changes to flashing.
The indication in the main display
changes to staying-on
and the
mode changes to the setting mode for the
operation start wait time.
4
Example: Stopping operation after 2 hours
and 30 minutes
Press the [
▼
] or
the [
▲
] key.
Press the [▼] or the [▲] key to change to
value in the main display to a wait time
before operation starts.
*
See step 5 of P.18
5
Press the
[ENTER] key.
Press the [ENTER] key.
The status changes to standy-by for the
program you have selected.
The main display indicates the present
measured temeperature.
The sub display indicates the remaining
time before starting operation.
The [STAND BY] lamp flashes to indicate
that the unit is in the stand-by mode.
The status of the [PROGRAM] lamp
changes from flashing to staying on.
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU
MEASURED TEMP
TROUBLE
REMOTE
FIXED- TEMP
PROGRAM
MODE
HEAT
STANDBY
ENTER
MENU