![Yamato BF200 Скачать руководство пользователя страница 7](http://html.mh-extra.com/html/yamato/bf200/bf200_manual_905412007.webp)
4
2. Before operating the Equipment
Precautions when installing the Equipment
1.
Choose proper place for installation
Do not install this Equipment in the place where:
・
the location is rough, dirty or un-leveled.
・
flammable gas, explosive gas or corrosive gas will be generated.
・
ambient temperature will be more than 35
℃
.
・
ambient temperature will fluctuate.
・
there is excessive humidity and dusty.
・
there is direct sunlight.
Install the Equipment(s) at the place with sufficient space.
2.
Install the Equipment on leveled location.
Install this Equipment on leveled floor. If it is installed on rough and/or slope floor, vibration
or noise will be occurred, and unexpected trouble and malfunction may be happened.
3.
Implement safety measures when installing the unit.
May be injured by moved and/or fallen this Equipment down by earthquake and/or
unexpected impact. Recommend to install this Equipment at the place away from the access
door and to take other safety steps.
4.
Must connect grounding wire properly.
・
Require to ground by Electrical Equipment Technical Standards Section 19-calss D in
Japan, if grounding terminal is not provided. Please contact with local dealer, local
electrician, or Yamato Customer Service Center.
・
Connect the terminals firmly to switch board of facilities or appropriate power plug.
Never connect grounding wire to gas line pipe, water line pipe or telephone grounding wire
due to fire or electric shock.
5.
Never disassembly nor modify the Equipment.
Never disassemble nor modify this Equipment. Those actions may cause this Equipment
malfunction, fire or electric shock.
6.
Never operate in an atmosphere where flammable or explosive gas is present.
Never operate this Equipment in an atmosphere where flammable or explosive gas is
present. This Equipment is not explosion-proof. Spark may be discharged by switching Earth
Le
akage Breaker (ELB) “ON(
|
)” and “OFF(○)” and also relay during operation, and then it
may cause fire or explosion.
See Chapter 19. “List of Dangerous Substances” for flammable and explosive gases on
page .74