4.3.2 Priority
4.4
Calibration
4.4.1 Overview
Why calibrate?
During operation, individual components of the Alyza IQ can age or become
dirty with time and thus change their characteristics, e.g.:
Flow through cell
LED of the photometer
Photo diodes
Color reagent
Regular automatic or manual calibration procedures help you to recognize any
changes of the measuring characteristics of the Alyza IQ.
2-point calibration
The slope and offset of the characteristic curve are adjusted with a 2-point cali-
bration.
Temperaturmodus
°C
°F
Temperature values are displayed in
the selected unit
Speichern u. Beenden
The Alyza IQ stores all changed set-
tings and the display switches to the
next higher level.
Beenden
The display switches to the next higher
level without storing the new settings.
Setting menu (general set-
tings)
Possible values
Description
The
Autom.Reinigung
,
Autom.Kalibrierung
and
Messintervall
set-
tings may result in the overlapping of the carrying out of different
functions.
In this case, the functions are carried out according to priority.
Function
Priority
Duration (min)
Autom.Reinigung
1
5 (1-channel variant)
Autom.Kalibrierung
2
40
Measuring
3
10