Maintenance and cleaning
99
ba76202d02
05/2020
5
Maintenance and cleaning
5.1
Hazard warnings
Read the chapter 2 Safety instructions,
22 before doing any maintenance
work. This is important for your personal safety.
NOTE
The interior of the measuring unit is temperature-controlled to 20 °C (68 °F).
With ambient temperatures over 25 °C (77 °F), condensation water may develop
on the cool surfaces and cause damage when the measuring unit is opened.
To avoid damage of the measuring unit due to the formation of condensation
water, always wait for the temperature adjustment (function "Öffnen der Mes-
seinheit vorbereiten") to be completed before opening the measuring unit.
Risks when
handling the
ChemBags
Notes on safely
handling the
ChemBags
If a ChemBag was damaged in the transport packaging, e.g. visibly leaked
liquid, do not use this ChemBag in the Alyza IQ.
Make sure you don't damage the ChemBags when unpacking them from the
transport packaging.
Do not use any sharp object when dealing with (or unpacking) the Chem-
Bags.
If possible, handle the ChemBags gripping only the cap in order not to kink
the ChemBags.
For work done with the locking device of the MultiPort valve open:
Prior to working with the locking device open make sure that the connected
ChemBags are safely suspended from the supporting rods, and that they are
not pressed or moved while the locking device is open.
Working with the ChemBags:
Prior to working with any connected ChemBags make sure that the locking
device of the MultiPort valve is closed.
WARNING
Dangerous chemicals.
Improper use of chemicals can cause damage to your
health.
Heed the following rules:
Read the labels of the ChemBags and follow the safety
instructions
Wear protective equipment (lab coat, protective gog-
gles, chemical resistant protective gloves, safety
shoes)