Alyza IQ NH4
Maintenance and cleaning
133
ba76202d02
05/2020
Opening the power
box
figure 5-17Mounting plate
1
Mounting plate
2
Fixing screw at the upper edge
3
Fixing nuts at the lower edge
4
Switch box with 24 VDC switch
9
Remove the mounting plate:
Lift the mounting plate upward over the threaded pins.
Tilt the upper edge of the mounting plate somewhat backwards and
remove the mounting plate from the housing downwards.
Place the mounting plate with the rear side down on a protected sur-
face (e.g. with cardboard).
2
1
4
3
Before opening:
ES: Antes de abrir:
10
Unscrew all nuts with safety disks (10 pieces) from the power supply
box and remove the lid of the power supply box.