Alyza IQ NH4
Lists
163
ba76202d02
05/2020
8.1.2 Informative messages
ES5552
Communication failure with ACS
* Check ACS connection
ES6552
Pressure on port X to high.
* Check the liquid circle for erros and replace maintenance parts if
necessary.
EI13Ex
EI1552
Operational voltage too low
* Check installation and cable lengths, Follow installation instruc-
tions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components, replace components
EI23Ex
EI2552
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instruc-
tions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components,
replace components
EI5552
The measuring unit is not compatible!
* Contact Service
EIA552
Communication fault between MIQ/Alyza and ACM
* Check cable connections
* Check the power supply of the Alyza IQ
* Contact service
Message code
Message text
Message code
Message text
II13Ex
II1552
Language not available,
Default language German
* Contact service
IC5552
Alyza IQ was successfully calibrated
* For calibration data, see calibration history