Maintenance and cleaning
121
ba76202d02
05/2020
Restarting
operation
5.7.3 Storing a used and cleaned filter plate
Proceed as follows to store the filter plate:
5.7.4 Replacing the filter plate of the sample filtration
NOTE
The filter plate is easily damaged. Never touch the filter plate with sharp-edged
objects or place any objects on it.
Maintenance
interval
As necessary, if cleaning does not help.
Proceed as follows to exchange the filter plate:
Dismounting the
filter plate
8
After the chemical cleaning, rinse the filter module with clean water.
Even a filter plate that is supposedly irreversibly blocked may be re-
stored by leaving it in the cleaning solution for a longer period of time
and then rinsing it several times.
9
After cleaning, check both sides of the filter plate for damage.
10
Remount the filter plate (see section 5.7.4 Replacing the filter plate of
the sample filtration,
121).
or
Store the cleaned filter plate (see section 5.7.3 Storing a used and
cleaned filter plate,
1
Clean the filter plate mechanically.
2
Clean the filter plate chemically.
3
Rinse the filter plate under flowing tap water.
4
To protect it against drying out, store the filter plate in the Filter-CL
cleaning container or in a sealed plastic bag.
Prior to each use, soak the filter plate in tap water.
If the filter plate has dried out, soak it in tap water for several hours.
1
Stop the Alyza IQ and activate the maintenance condition at the
IQ S
ENSOR
N
ET
.