Alyza IQ NH4
Commissioning
65
ba76202d02
05/2020
The filter module (
FM/PC
) and the slide must be completely submersed
(max. 40 cm).
Take changing water levels into account when mounting the filter membrane
module.
The lower edge of the filter module must be mounted with at least 10 cm dis-
tance to the bottom.
NOTE
Dirt particles and contamination in the sample tube can block the valves of the
measuring unit.
During the cleaning activities, make sure that no dirt particles get into the open
tube ends or the connector of the filter module.
Protect the open tubes and connectors during the cleaning activities, e.g. with
blind plugs.
Installation
In special cases (e.g. in a channel) it is better to mount the FM/PC
filter module in a horizontal position in the flow direction.
An adapter for horizontal mounting is available as an accessory.
1
Mount the rail of the basin holder for filtration in the basin.
If necessary, extend the rail with the M-EXT 1.5 extension accessory.
figure 3-27Sample filtration device (installed)
1
Chain (scope of delivery: basin holder for filtration M 1.5)
2
Guide rail (scope of delivery: basin holder for filtration M 1.5)
3
Height adjustable slide (
scope of delivery: Suction line)
4
Intake line (
scope of delivery: Suction line)
5
Sleeve tube (scope of delivery: Suction line)
6
Filter module (
FM/PC
) with filter plate (
Filter/PC
)
2
Mount the filter module on the slide of the basin holder for filtration.
3
Connect the intake line to the filter module.
2
1
3
6
5
4