Alyza IQ NH4
What to do if ...
151
ba76202d02
05/2020
Status
FEHLER
Sample intake does not work
Check the filtration pump
Check and, if necessary, clean the filter
plate
Switch on the filtration pump
Clean the overflow vessel and intake
line
Status
FEHLER
The temperature in the mea-
suring unit or photometer unit
Alyza IQ is outside the allowed
range
(e.g. due to the air circulation
in the Alyza IQ being
impeded)
Measuring (automatically or manually) is
only possible if the operating temperature
in both the measuring unit and photometer
unit is in the allowed range.
Check for contamination and, if neces-
sary, exchange the filter mats at the
housing (see section 5.8.2,
Check for contamination and, if neces-
sary, clean the bug screen
Check for contamination and, if neces-
sary, clean the condensate drain
adapter
Status
FEHLER
Danger of condensation water
forming in the measuring unit
View the details of the error
e.g. in the Alyza menu / tab
Status
(dis-
play details with
<OK>
)
or
View the log book message
Start the dehumidifying program for the
measuring unit
Alyza menu/ tab
Wartung
/
Wartung Mes-
seinheit
/
Messeinheit entfeuchten
Alyza IQ STARTEN
Power failure
In the Alyza menu (tab
Wartung
), start
the Alyza IQ.
If necessary, activate the function "Auto-
matic start after power failure“
Unknown
See log book or Alyza menu / tab
Status
(display details with
<OK>
)
Cause
Remedy