What to do if ...
Alyza IQ NH4
150
ba76202d02
05/2020
6
What to do if ...
Display “----”
Information on the cause of the indication is in the log book and in
the Alyza menu / tab
Status
.
Cause
Remedy
No (valid) measured value
available
Start measurement and wait for the mea-
surement to be finished (10 min)
Three erroneous automatic
calibrations in succession
Check the calibration standards
(e.g. expiry date)
Check the connection of the calibration
standards
Refill the tubes for the ChemBags with
standard solutions
Alyza menu / tab
Wartung
/
Manuelle
Funktionen
/
Befüllen (einzeln)
Carry out a cleaning cycle
Alyza menu / tab
Wartung
/
Manuelle
Funktionen
/
Reinigen
Exchange calibration standards
Contact the service department
Status
FEHLER
The Alyza IQ is stopped
General proceeding
View the details of the error
e.g. in the Alyza menu / tab
Status
(dis-
play details with
<OK>
)
or
View the log book message
Reset the error:
(Alyza menu, tab
Wartung
/
Manuelle
Funktionen
/
Fehler zurücksetzen
)
Remedy the error
Alyza IQ STARTEN
If the
Status
FEHLER
is still there:
Contact the service department
Status
FEHLER
The liquids from one or sev-
eral ChemBags are not trans-
ported to the MultiPort valve
View the log book message
Check the filling level of the ChemBags
Check the connection of the tubes
Empty and refill the tube
Alyza menu / tab
Wartung
/
Manuelle
Funktionen
/
Entleeren
and
Befüllen