Measurement / Operation
Alyza IQ NH4
82
ba76202d02
05/2020
Messbereich
With the setting
Messmodus
:
NH4-N
Depending on the
Messmodus
setting,
different measuring ranges can be
selected.
For each measured parameter, a small
and a large measuring range can be
selected.
To change between the large and small
measuring range, different ChemBags
have to be installed (see section 5.6
Changing the measuring range,
0,02 .. 5,00 mg/L
0,10 .. 20,00 mg/L
0,02 .. 5,00 ppm
0,10 .. 20,00 ppm
With the setting
Messmodus
:
NH4
0,03 .. 6,00 mg/L
0,13 .. 25,00 mg/L
0,03 .. 6,00 ppm
0,13 .. 25,00 ppm
With the setting
Messmodus
:
NH3-N
0,02 .. 5,00 mg/L
0,10 .. 20,00 mg/L
0,02 .. 5,00 ppm
0,10 .. 20,00 ppm
Offset-Korrektur
With the setting
Messmodus
NH4-N
or
NH3-N
and
Messbereich
0,02 .. 5,00
mg/L
(small MR)
- 0.50…
0.00
…+ 0.50
mg/L
For other measuring modes and mea-
suring ranges, the value ranges are
adjusted.
With the setting
Messmodus
NH4-N
or
NH3-N
and
Messbereich
0,10 ..
20,00 mg/L
(large MR)
- 2.00…
0.00
…+ 2.00
mg/L
Setting menu
Possible values
Description
Setting menu (general set-
tings)
Possible values
Description
Messintervall
10
, 15, 20, 30, 45,
60 min
2, 4, 6, 8, 12 h
Measuring interval
Dämpfung
Aus
Signalglättung
Medianfilter
Methods for filtering the signals
A signal filter in the sensor reduces the
limits of variation of the measured
value.
With the setting
Signalglättung
:
The signal filter is essentially charac-
terized by the response time t90.
Response time of the signal filter (in
seconds).
This is the time after which 90 % of a
signal change is displayed.
Ansprechzeit t90
100
, 200, 300, 400 s