Multi 3430
Operation
60
ba75865e19
04/2015
the sensor evaluation with the sensor symbol.
The QSC initial calibration is completed. The sensor is calibrated. If you
want to calibrate with special buffers for your measurements, you can
subsequently carry out a normal calibration with your buffers. The ref-
erence values determined with the QSC calibration are also used for
the evaluation of normal calibrations. In the measured value display,
the color scale of the QSC function is always displayed. A double arrow
on the color scale indicates the current sensor evaluation.
You can carry out QSC control calibrations at greater intervals than
normal calibrations.
A QSC control calibration can, e.g. be useful if the sensor evaluation
noticeably changed (after some normal calibrations).
Carrying out a
QSC
control calibration
4
As soon as the three-point calibration has been successfully
carried out you can decide whether to accept or discard the cal-
ibration as the QSC initial calibration.
15.03.2014 08:00
Info
QSC color scale
The double arrow on
the color scale indi-
cates the current sen-
sor evaluation.
1
Open the menu for measurement settings with
<MENU/
ENTER>
.
2
In the
QSC
menu, select
Control calibration
with
<
><
>
.
The calibration display appears. AutoCal QSC-Kit is displayed
as the buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use
other buffers, you will have no valid QSC control calibration.
3
Follow the user guide.
The calibration is carried out like a normal three-point calibra-
tion. As soon as the three-point calibration has been success-
fully carried out you can decide whether to accept or discard the
calibration as the QSC control calibration.
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......