Multi 3430
Operation
56
ba75865e19
04/2015
4.7.8
Displaying calibration records
The calibration data can be displayed and then output to the interface.
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item,
Calibration
/
Calibration record
. To open it in the measured
value display, press the
<CAL
_
>
key.
The calibration records of the last 10 calibrations are to be found in the
menu,
Calibration
/
Calibration data storage
/
Display
. To open the
Cal-
ibration
menu, press the
<MENU/ENTER>
key in the measured value
display.
Menu item
Setting/
function
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration
records.
Further options:
Scroll through the calibra-
tion records with
<
><
>
.
Output the displayed cali-
bration record to the inter-
face with
<PRT>
.
Output all calibration
records to the interface
with
<PRT
_
>
.
Quit the display with
<ESC>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
flash drive or printer
-
Outputs the calibration data
storage to the interface USB-
B (USB flash drive/USB
printer)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data
storage to the interface USB-
B (PC)
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......