What to do if...
Multi 3430
121
ba75865e19
04/2015
No stable measured
value
Obviously incorrect
measured values
6.3
Dissolved oxygen
Error message,
OFL
Cause
Remedy
IDS pH sensor:
– Junction contaminated
– Clean junction
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Adjust temperature if neces-
sary
IDS pH test sample:
– Conductivity too low
– Use suitable IDS pH sensor
– Temperature too high
– Use suitable IDS pH sensor
– Organic liquids
– Use suitable IDS pH sensor
Cause
Remedy
IDS pH sensor:
– IDS pH sensor unsuitable
– Use suitable IDS sensor
– Temperature difference
between buffer and test sample
too great
– Adjust temperature of buffer
or sample solutions
– Measurement procedure not
suitable
– Follow special procedure
Cause
Remedy
– Measured value outside the
measuring range
– Select a suitable measuring
medium