Multi 3430
Operation
50
ba75865e19
04/2015
Continuing with three-
to five-point calibration
13
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with
<MENU/ENTER>
.
The calibration display for the next buffer appears (voltage dis-
play).
14
If necessary, finish the calibration procedure as a two-point cal-
ibration with
<M>
.
The calibration record is displayed.
15.03.2014 08:00
15
Thoroughly rinse the IDS pH sensor with deionized water.
16
Immerse the IDS pH sensor in the next buffer solution.
17
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
><
>
.
18
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
15.03.2014 08:00