4.10.3 Temperature compensation
The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
Tr20
or
Tr25
.
You can select one of the following temperature compensation meth-
ods:
Nonlinear temperature compensation
(
nLF
) according to EN 27
888
Linear temperature compensation
(Lin)
with adjustable coeffi-
cients of 0.000 ... 10.000 %
/
K
No temperature compensation (off)
Note
The reference temperature and temperature compensation are set in
the menu for the parameter, conductivity (see section 4.10.4).
Application tips
Select the following temperature compensations given in the table
according to the respective test sample:
Test sample
Temperature compensation
Display
indicator
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water
nLF
according to EN 27 888
nLF
Other aqueous solu-
tions
lin
Set linear
temperature coefficient
0.001 ... 10.000 %/K
lin
Salinity (seawater)
Automatic
nLF
according to
IOT (International Oceanogra-
phic Tables
Sal,
nLF
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......