Operation
Multi 3430
81
ba75865e19
04/2015
Type
Measuring cell used
Cal
Measuring cells whose cell con-
stant is determined by calibration
in the KCL control standard solu-
tion.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant
is displayed in the status line.
man
Freely adjustable cell constant in
the range 0.450 ... 0.500 cm
-1
.
Man. cell const.
0.450 ...
0.475
...
0.500 cm
-1
The cell constant and be dis-
played and set here
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature com-
pensation (see section 4.10.3).
This setting is only available for
the measured parameters,
ϰ
and
ρ
.
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
10.000 %/K
Coefficient of the linear tempera-
ture compensation.
This setting is only available
when the linear temperature
compensation is set.
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for
the measured parameters,
ϰ
and
ρ
.
TDS factor
0.40 ...
1.00
Factor for TDS value
Reset
-
Resets all sensor settings to the
delivery condition (see section
4.16.1)
Menu item
Possible
setting
Description
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......