Multi 3430
Operation
104
ba75865e19
04/2015
Settings
You can configure the automatic storage function with the following set-
tings:
Starting the automatic
storage function
To start the automatic storage function, select
continue
with
<
><
>
and confirm with
<MENU/ENTER>
. The meter switches to the mea-
sured value display.
ID number
1
Interval 30
s
Duration 180
min
continue
0d03h00min
0
1d17h33min
Automatic data storage
15.03.2014 08:00
Set entire storing dura-
tion
Max. available storing
duration
Graphical display of the
storage usage
Menu item
Possible
setting
Description
ID number
1 ... 10000
ID number for the dataset series
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min,
10 min,
15 min,
30 min,
60 min
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.
Duration
1 min ... x
min
Storage duration.
Specifies after which time the auto-
matic storage should be terminated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......